Antaris EX-IP Specifications
Electrical
All electrical signals enter the enclosure through connectors rated to
maintain the enclosure Ingress Protection rating. All wiring must meet
local codes.
If the system does not include an internal computer, an Ethernet cable is
used to connect the computer to the spectrometer.
Plumbing and pipe fitting
Each system requires fiber optic cabling for sample data collection, and
clean, dried nitrogen or air for purge. Depending on your application,
optional Ethernet (for remote computer and spectrometer control) and
analog and digital PLC transmission wiring also may be used.
Item
Specification
Enclosure purge enclosure: 0.25-inch NPT female
supply line: 0.25-inch NPT male
Fiber optic
spectrometer: SMA 905
cable: SMA 905 nut connector
1 meter tail required for spectrometer connection and serviceability.
Label with connection and sampling location.
98
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific