Installation Process
Calibrating the relief valve
A spark arrestor relief valve protects the enclosure from over pressurization.
Use the following steps to confirm that the valve is calibrated properly.
1.
Use the key and open the Mini Purge compartment.
If you are unfamiliar with opening the Mini Purge compartment, see
the documentation that came with the enclosure.
Key
2.
Press down on the quick connect collar and remove the red plug on
the low pressure sensor.
This opens the system pressure test point.
Quick
connect
collar
Plug
56
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific