Installation Process
Opening
the enclosure
Specific information about component cooling cycles and times is included
in documentation that accompanied the enclosure and on labels on the
enclosure itself. Use these instructions whenever you wish to open the
enclosure.
Danger
Avoid explosion and fire hazard. The NIR source is very hot. In volatile
environments always wait at least the amount of time specified on the
labeling on the enclosure before opening the enclosure.
This allows internal components to cool to a safe level.
To open the enclosure:
1.
Remove the power from the branch circuit servicing the
instrument.
2.
Wait at least at least the specified cooling time before opening the
enclosure to allow internal parts to cool below the flash point of the
surrounding atmosphere.
Cooling time is listed on a label on the outside of the enclosure.
72
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific