Antaris EX-Z1 Specifications
Shelf or
table top mounting
Optionally, a system may be mounted to a table top or shelf. The surface
must be capable of supporting 274 kg (600 lb), or two times the weight of
the instrument. See the section entitled “Lifting or moving” in the “Safety”
chapter earlier in this manual for lifting/moving precautions.
In this configuration, the wall mount flanges may be removed.
The equipment must be on a surface that is within 2 degrees of level for
proper operation.
For anchoring the enclosure, we recommend using high-quality steel
10 mm (3/8 in) bolts that are capable of withstanding the following forces:
Bolt characteristic
Specification
Shear force
250 N (56 lb)
Tensile force
300 N (68 lb)
Pull-out force
30 N (68 lb)
Thermo Fisher Scientific
Antaris EX Safety, Installation, and Site Preparation Guide