Installation Process
Danger
Avoid explosion and fire hazard. The NIR source is very hot. In volatile
environments always wait at least the amount of time specified on the
labeling on the enclosure before opening the enclosure.
This allows internal components to cool to a safe level.
This procedure requires the following tools:
•
Service override switch box
•
1/8-inch flat-blade screwdriver
Use the procedure that follows to override the protections for service and
then return the analyzer to normal operating mode.
1.
Confirm that the atmosphere is non-hazardous.
2.
Loosen the set screw on the MiniPurge interface unit (MIU) and
then remove the cover.
Loosen
screw
Locate
screw
Remove cover
3.
Locate the Service Override connector inside the MIU and connect
the service override switch box.
Thermo Fisher Scientific
Antaris EX Safety, Installation, and Site Preparation Guide
75