Installation Process
Caution
Avoid tip and crush hazards. Once the enclosure is placed on the cart,
secure it with bolts before moving the cart. Never move the cart unless all 4
anchor bolts are in place and tightened completely.
2.
Place the instrument onto the stand with the door facing the open
side of the cart.
Align the holes in the stand surface with those in the bottom of the
enclosure.
3.
Insert bolts through the stand into the bottom of the enclosure and
tighten a few turns.
4.
Once all 4 bolts are inserted into the enclosure, tighten them all to
securely fasten the enclosure to the stand.
For a table top or shelf installation:
1.
Drill the table top or shelf for anchoring bolts.
Check the schedule drawings for your instrument for drill size and
placement. Check the installation and regulatory compliance
specifications for your instrument (later in this manual) for anchoring
bolt specifications.
2.
Place the instrument on the table top or shelf, oriented so that the
door is accessible and there is adequate clearance to open the door
completely.
Align the holes in the bottom of the enclosure with those in the table
top or shelf.
36
Antaris EX Safety, Installation, and Site Preparation Guide
Thermo Fisher Scientific