PRISM Field Upgrade Kit Instructions
26.
Reinstall the rear panel hardware. Tighten the seven BNC nuts to 0.9 N·m
(7.97 in-lb).
27.
Place the cover over the instrument and replace the top cover screws. Tighten
the cover screws to 0.9 N·m (7.97 in-lb).
28.
Apply the MPX-UP serial number label below the existing MPX serial
number label on the bottom of the instrument.
NOTE.
The MPX-UP serial number label is included in the upgrade kit. The label
lets Factory Service know that the instrument contains the upgrade kit.
Figure 28: Installing the MPX-UP serial number label
CAUTION.
The MPX-UP label is not inset and is at risk of damage. To avoid
damaging the label, use caution when inserting or removing the instrument.
29.
Upgrade the instrument firmware.
30.
31.