PRISM Field Upgrade Kit Instructions
17
17.
Locate the new Eye mezzanine board from the kit (part number 878-
1285-XX).
18.
Line up the Eye mezzanine board with both of the connectors and press down
until it is fully engaged as show in the following figure.
CAUTION.
Bending the Eye mezzanine board can damage the board or
instrument. Avoid fully inserting the connectors until they are properly aligned;
this prevents the bending of the board.
Figure 23: Install the 12G Eye mezzanine board on the main board
19.
Use the TORX-driver with a T-10 TORX tip to secure the Eye mezzamine
board to the main board using the two screws provided in the kit. Tighten
these screws to 0.6 N·m (5.31 in-lb).
Figure 24: Attach the Eye mezzanine board