BASIC SETTING
45
Default option
Off
Steps
10.
Press down arrow key to move
cursor to
Peak Over
field for V-
Range and I-Range, respectively.
11.
Use soft keys to select and confirm
the desired option. Press the
More
soft key to toggle among pages for
Peak Over of V-Range and I-Range.
Option
When the occurrence of peak over-range happens
in Auto range mode, user is able to define a
measurement range to switch to. The available
options for each mode are listed below.
When it is under CF3 mode for V-Range.
Off/15V/30V /60V/150V/300V/600V
When it is under CF6/6A mode for V-Range.
Off/7.5V/15V /30V/75V/150V/300V
When it is under CF3 mode for I-Range.
Off/5mA/10mA/20mA/50mA/100mA/200mA/0.5A
/1A/2A/5A/10A/20A
When it is under CF6/6A mode for I-Range.
Off/2.5mA/5mA/10mA/25mA/50mA/100mA/250
mA/0.5A/1A/2A/5A/10A
Default option
Off