Input
Item
Specifications
Input type
Voltage Floating input through resistive voltage
divider
Current
Floating input through shunt
Measure range
Voltage
15 V, 30 V, 60 V,150 V, 300 V, 600 V
Current
Direct
input
5 mA, 10 mA, 20 mA, 50 mA, 100 mA,
200 mA, 0.5 A, 1 A, 2 A, 5 A, 10 A, 20 A
Sensor
input
EX1: 2.5 V, 5 V, 10 V
EX2: 50 mV, 100 mV, 200 mV, 500 mV, 1
V, 2 V
Input impedance
Voltage
Input resistance: approach 2 M
Ω
Current
Direct input range 5 mA
~ 200 mA
Input resistance:
approach 505 m
Ω
Direct input range 0.5A
~ 20 A
Input resistance:
approach 5 m
Ω
Sensor input
Input range 2.5 V ~ 10 V
(EX1)
Input resistance:
approach 100 k
Ω
Input range 50 mV ~ 2 V
(EX2)
Input resistance:
approach 20 k
Ω
Continuous maximum
allowable input
Voltage
peak value of 1.5 kV or RMS value of 1
kV, whichever is less
Current
Direct input range 5 mA
~ 200 mA
peak value of 30 A or
RMS value of 20 A,
whichever is less
Direct input range 0.5A
~ 20 A
peak value of 100 A or
RMS value of 30 A,
whichever is less
Sensor
input
peak value less than or equal to 5 times
of the rated range
Input bandwidth
DC, 0.1 Hz ~ 100kHz
Continuous maximum
Common-mode voltage
600 Vrms, CAT
Ⅱ
Line filter
select OFF or ON (cut off frequency of 500 Hz)
Frequency filter
select OFF or ON (cut off frequency of 500 Hz)
A/D converter
Simultaneous conversion voltage and current inputs
Resolution
16bits
Maximum conversion rate
Approx. 300kHz