Parameter/
Return parameter
<x>
<Function>
<Element>
1 to 10 (display).
U|UPPeak|UMPeak|I|IPPeak|IMPeak
|P|PPPeak|PMPeak|S|Q|LAMBda|CFU
|CFI|PHI|FU|FI|UTHD|ITHD|MATH|MCR
1(If <Element> is omitted, the element
is
set to 1)( For the T3PM1100, only set to
1
or omitted)
Example
:DISPLAY:NORMAL:ITEM1 U,1
:DISPLAY:NORMAL:ITEM1?
->:DISPLAY:NORMAL:ITEM1 U,1
<Function>
Function
T3PM1100
Indicator
U
Voltage V
[V]
UPPeak
Maximum voltage: V+pk
[V+pk]
UMPeak
Minimum voltage: V-pk
[V-pk]
I
Current I
[I]
IPPeak
Maximum current: I+pk
[I+pk]
IMPeak
Minimum current: I-pk
[I-pk]
P
Active power P
[P]
PPPeak
Maximum power: P+pk
[P+pk]
PMPeak
Minimum power: P-pk
[P-pk]
S
Apparent power S
[VA]
Q
Reactive power Q
[VAR]
LAMBda
Power factor
λ
[PF]
CFU
Voltage factor
λ
[CFV]
CFI
Current factor
λ
[CFI]
PHI
Phase difference
Φ
[DEG]
FU
Voltage frequency fV
[VHz]
FI
Current frequency fI
[AHz]