APPENDIX
205
DC
Measurement
synchronization source
Select voltage, current, or off
In the case of Auto Update Rate, select the voltage or
current from the equipped element.
Line filter
Select OFF or ON (cutoff frequency at 500 Hz).
Peak measurement
Measures the peak (max, min) value of voltage,
current or power from the instantaneous voltage,
instantaneous current or instantaneous power that is
sampled.
Zero-level compensation
Removes the internal offset of the measure unit
(After measurement range is changed)
Measurement parameters
Voltage
Vrms , Vmn, Vdc , Vac
Current
Irms , Idc , Iac
Active Power
P
Apparent Power
VA
Reactive power
VAR
Power Factor
PF
Crest Factor
CFI,CFV
Phase Angle
DEG
Frequency
IHz and VHz
Voltage Peak
V+pk and V-pk
Current Peak
I+pk and I-pk
Active Power Peak
P+pk and P-pk
Total Harmonic Distortion
THDI and THDV
Mathematical Computation MATH
Maximum Current Ratio
MCR
Frequency Measurement
Item
Specifications
Measurement item
Voltage and current
Measurement frequency
range
Data update interval Measurement Frequency Range
0.1 s
20 Hz
≤
f
≤
100 kHz
0.25 s
10 Hz
≤
f
≤
100 kHz
0.5 s
5 Hz
≤
f
≤
100 kHz
1 s
2.0 Hz
≤
f
≤
100 kHz
2 s
1.0 Hz
≤
f
≤
100 kHz
5 s
0.5 Hz
≤
f
≤
100 kHz
10 s
0.2 Hz
≤
f
≤
100 kHz
Summary of Contents for Everywhereyoulook T3PM1100