COMMAND OVERVIEW
129
Example
:AOUTPUT:NORMAL:CHANNEL1 NONE
Turns D/A channel1 output off (0V)
:AOUTPUT:NORMAL:CHANNEL1?
->:AOUTPUT:NORMAL:CHANNEL1 I,1
:AOUTput[:NORMal]:IRTime
Description
Sets or returns the rated integration time that is
used in the D/A output of the integrated value.
Syntax
:AOUTput[:NORMal]:IRTime {<NRf>,<NRf>,<NRf>}
Query Syntax
:AOUTput[:NORMal]:IRTime?
Parameter/
Return parameter
1st <NRf>
2nd <NRf>
3rd <NRf>
0~9999(hour)
0~59(minute)
0~59(second)
Example
:AOUTPUT:NORMAL:IRTIME 1,1,10
:AOUTPUT:NORMAL:IRTIME?
->:AOUTPUT:NORMAL:IRTIME 1,1,10
:AOUTput[:NORMal]:MODE<x>
Description
Sets or returns a D/A range mode. (Remote
Control Only)
Syntax
:AOUTput[:NORMal]:MODE<x>
{FIXed|MANual|COMPare}
Query Syntax
:AOUTput[:NORMal]:MODE<x>?
Parameter/
Return parameter
<x>
FIXed
MANual
COMPare
1 to 4 (channel)
Fixed range mode.
Manual range mode.
Comparator mode.
Set
Query
Set
Query