BASIC SETTING
27
2.
Use up and down arrow keys to
select the desired range.
3.
Press
Enter
button to confirm your
selection.
Enter
Available range
Crest Factor
is 3:
AUTO, 5mA, 10mA, 20mA, 50mA,
100mA, 200mA, 0.5A, 1A, 2A, 5A, 10A,
20A
Crest Factor
is 6/6A:
AUTO, 2.5mA, 5mA, 10mA, 25mA,
50mA, 100mA, 250mA, 0.5A, 1A, 2.5A,
5A, 10A
Note
When it is under the manual setting ranging from
5mA to 200mA, the manual range adjusts to auto
range automatically if input current exceeds 700mA.
Note
When the measurement range is set manually, if the
range status icon lights in green, the measured value
meets the setting range. On the contrary, if the range
status icon lights in red, the measured value doesn’t
meet the best setting range. In this case, it is better to
switch to another range to get more accurate
measurement results.
Note
The P.I status icon lights in red when the current
measurement circuit detects that the measured value
exceeds setting range by 3 folds (CF is set to 3) or 6
folds (CF is set to 6/6A).