T3PM1100 User Manual
40
6.
Use soft keys to increase or
decrease the interval.
Option
The setting range for Interval is from 00:00:00 to
99:59:59.
Default value
00:00:00
Note
When it is set 00:00:00, the interval for measure
storage will be synchronized with the designated
Data Update Rate.
Storage stops in the following circumstances:
When data has been stored to all blocks,
Normal measure data can be stored 10000
blocks and Normal with Harmonic data can be
stored 1000 blocks.
When the storage setting is set to Off (while
storage is in progress)
If you press the HOLD key to hold the display
while storage is in progress, the measurement
operation and the storage interval time counter
are held (paused), which causes the storage
operation itself to be held. If integration is in
progress, this instrument continues
measurement and integration in the
background.