T3PM1100 User Manual
154
Return parameter 0
Turn the scaling function off.
1
Turn the scaling function on.
Example
:INPUT:SCALING:STATE ON
:INPUT:SCALING:STATE?
->:INPUT:SCALING:STATE 1
[:INPut]:SCALing:{VT|CT|SFACtor}:ELEMen
t<x>
Description
Sets or returns the VT ratio, CT ratio, or power
coefficient of the specified element.
Syntax
[:INPut]:SCALing:{VT|CT|SFACtor}:ELEMent<x>
{<NRf>}
Query Syntax
[:INPut]:SCALing:{VT|CT|SFACtor}:ELEMent<x>?
Parameter/
Return parameter
<x>
1(If <Element> is omitted, the element
is
set to 1)( For the T3PM1100, only set to
1
or omitted)
<NRf>
0.001 to 9999
Example
:INPUT:SCALIG:VT:SRATIO:ELEMENT1 10
:INPUT:SCALIG:VT:SRATIO:ELEMENT1?
->:INPUT:SCALIG:VT:SRATIO:ELEMENT1 10
[:INPut]:SYNChronize
Description
Sets or returns the synchronization source.
Syntax
[:INPut]:SYNChronize {VOLTage|CURRent|OFF}
Query Syntax
[:INPut]:SYNChronize?
Parameter/
Return parameter
VOLTage
Select the voltage synchronization
source.
Set
Query
Set
Query