T3PM1100 User Manual
206
20 s
0.1 Hz
≤
f
≤
100 kHz
Auto ( * )
0.1 Hz
≤
f
≤
100 kHz
( * ) Limit of the measurement lower limit frequency
by the Timeout setting
Timeout
lower limit frequency
1 s
2.0 Hz
5 s
0.5 Hz
10 s
0.2 Hz
20 s
0.1 Hz
Measurement range
Auto switching among six types: 100mHz, 1 Hz, 10
Hz, 100 Hz, 1 kHz, 10 kHz, and 100 kHz.
Frequency filter
Select OFF or ON (cut off frequency of 500 Hz)
Accuracy
Requirements
When the input signal level is
30% or more of the
measurement range If the
crest factor is set to 3.
(60% or more if the crest
factor is set to 6 or 6A)
• Frequency filter is ON when
measuring voltage or current
of 200 Hz or less.
± (0.06% of reading)
Integration
Item
Specifications
Mode
Select manual integration mode, standard integration
mode, or repetitive integration mode.
Timer
Automatically stop integration by setting a timer.
Selectable range: 0 hours 00 minutes 00 seconds to
9999 hours 59 minutes 59 seconds
Accuracy
±(Power accuracy (or current accuracy) + 0.1% of
reading) (fixed range)
Range setting
Auto range or fixed range is available for Integration
Timer accuracy
±0.02%
Remote control
Start, stop and reset operations are available using
an external remote signal. (option)