T3PM1100 User Manual
20
GND Terminal
Connects the GND (ground)
terminal to the earth ground.
Voltage/Current
input terminal
I
I
20A MAX
CAT II 600V
CURRENT
V
VOLTAGE
V
600V MAX
EXT1
EXT2
GPIB
10V MAX
2V MAX
Voltage/Current
input terminal is
used to connect
the main
measurement
signals.
External Input 1/2
EXT1
EXT2
10V MAX
2V MAX
Connects output signal to the EXT1
terminal which receives up to 10V, or
the EXT2 terminal that receives at the
maximum of 2V. See page 53 for setting.
Warning
Do not use damaged device. Before using the
equipment, check its housing first to sure ensure
there are no cracks. Do not operate this device in
an environment containing explosive gases, steam
or dust.
The maximum measurable current and voltage are
600 V and 20A for voltage and current terminals of
the rear panel of the T3PM1100. Do not input
exceeded voltage and current, otherwise it will burn
the device.
The maximum input voltage are 10 V and 2V for
EXT1 and EX2 terminals of the rear panel of the
T3PM1100. Do not input exceeded voltage,
otherwise it will burn the device.
Always use the supplied cable for connection.
Before connecting the device, observe all the safety
symbols marked on the device.
Turn off the power to the device and the application
system before connecting I/O terminals.
Do not install replacement parts on the device or
perform any unauthorized modifications.