T3PM1100 User Manual
176
:NUMeric:LIST:VALue
Description
Returns the harmonic measurement numeric list
data.
Query Syntax
:NUMeric:LIST:VALue? {<NRf>}
Parameter
<NRf>
1 to 8 (item number)
Example
• if <NRf> is specified :NUMERIC:LIST:VALUE? 1
-> 103.58E+00,NAN,103.53E+00,0.09E+00,2.07E+00,
0.04E+00,.. (omitted)..,0.01E+00,0.01E+00
(up to 52 data values)
• if <NRf> is omitted (when :NUMeric:LIST:NUMber
is set to 5)
:NUMERIC:LIST:VALUE?
-> 103.58E+00,NAN,103.53E+00,0.09E+00,2.07E+00,
0.04E+00,..(omitted)..,0.00E+00,0.00E+00
(up to 52*5 = 260 data values)
• When :NUMeric:FORMat is set to {FLOat}
:NUMERIC:LIST:VALUE?
-> #N (N-digit byte number)(data byte sequence)
Note
A single numeric list data item consists of up to
52 items of numeric data in the following order:
TOTal, DC, 1st
harmonic, ..., :NUMeric:LIST:ORDer.
If <NRf> is specified, only the numeric list data of
the specified item number is output (up to 52
items of data)
If <NRf> is omitted, the numeric list data of item
numbers from 1 to :NUMeric:LIST:NUMber is
output in order (up to 52 times the number
specified by :NUMeric:LIST:ORDer)
Query
Summary of Contents for Everywhereyoulook T3PM1100
Page 1: ...T3PM1100 Digital Power Meter User Manual ...
Page 191: ...COMMAND OVERVIEW 191 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Page 209: ...APPENDIX 209 Status system The diagram below is a description of the status system ...
Page 213: ...APPENDIX 213 Dimensions Units mm ...