
COMMAND OVERVIEW
169
Parameter/
Return parameter
<x>
1 to 50 (item number)
NONE
No output item.
<Function>
{U|UPPeak|UMPeak|I|IPPeak|IMPeak
|P|PPPeak|PMPeak|S|Q|LAMBda|CFU
|CFI|PHI|FU|FI|UTHD|ITHD|WH
|WHP|WHM|AH|AHP|AHM|TIME
|URANge|IRANge|MATH|MCR}
{UK|IK|PK|LAMBDAK|PHIK|PHIUK|P
HIIK|UHDFK| IHDFK|PHDFK}
<Element>
1(If <Element> is omitted, the
element is set to 1)( For the
T3PM1100, only set to 1 or omitted)
<Order>
{TOTal|DC|<NRf>} (<NRf> = 1 to 50)
Example
:NUMERIC:NORMAL:ITEM1 U,1
:NUMERIC:NORMAL:ITEM1?
->:NUMERIC:NORMAL:ITEM1 U,1
:NUMERIC:NORMAL:ITEM1 UK,1,1
:NUMERIC:NORMAL:ITEM1?
->:NUMERIC:NORMAL:ITEM1 UK,1,1
If <Order> is omitted, the order is set to TOTal.
This instrument does not measure data for
<Order> = DC.
<Function>
Function
T3PM1100 Indicator
U
Voltage V
[V]
UPPeak
Maximum voltage: V+pk
[V+pk]
UMPeak
Minimum voltage: V-pk
[V-pk]
I
Current I
[I]
IPPeak
Maximum current: I+pk
[I+pk]
IMPeak
Minimum current: I-pk
[I-pk]
P
Active power P
[P]
PPPeak
Maximum power: P+pk
[P+pk]
Summary of Contents for Everywhereyoulook T3PM1100
Page 1: ...T3PM1100 Digital Power Meter User Manual ...
Page 191: ...COMMAND OVERVIEW 191 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Page 209: ...APPENDIX 209 Status system The diagram below is a description of the status system ...
Page 213: ...APPENDIX 213 Dimensions Units mm ...