Crest factor 3
Vrms or Irms exceeds 130% of the
currently set measurement range.
Vpk, Ipk value of the input signal
exceeds 300% of the currently set
measurement range.
Crest factor 6
Vrms or Irms exceeds 130% of the
currently set measurement range.
Vpk, Ipk value of the input signal
exceeds 600% of the currently set
measurement range.
Crest factor 6A
Vrms or Irms exceeds 260% of the
currently set measurement range.
Vpk, Ipk value of the input signal
exceeds 600% of the currently set
measurement range.
Auto-range decline
The range is downed when all of the following
conditions are met.
Crest factor 3
Vrms or Irms is less than or
equal to 30% of the
measurement range.
Vrms or Irms is less than or
equal to 125% of the next
lower measurement range.
Vpk, Ipk value of the input
signal exceeds 300% of the
currently set measurement
range.
Crest factor 6 or 6A
Vrms or Irms is less than or
equal to 30% of the
measurement range.
Vrms or Irms is less than or
equal to 125% of the next
lower measurement range.
Vpk, Ipk value of the input
signal exceeds 600% of the
currently set measurement
range.
Display mode
Switching
Vrms (the true RMS value of voltage and current)
VOLTAGE MEAN (the rectified mean value
calibrated to the RMS value of the voltage)
AC