REMOTE CONTROL
111
Option
9600, 19200, 38400, 57600,115200
Default value
9600
6.
Press down arrow key to move
cursor to
Terminator
field.
7.
Use soft keys to select and confirm
the
Terminator
option.
Option
CR, LF, CR+LF
Terminator indicates the end
of line for return message.
Default value
CR+LF
RS232
Configuration
Selectable Baud rate
9600, 19200, 38400,
57600, 115200
Parity
None
Hardware flow control
Off
Data Bits
8
Stop bit
1
RS232 Pin
Assignments
Pin 2: RxD
Pin 3: TxD
Pin 5: GND
Pin 1, 4, 6 ~ 9: No
Connection
1 2 3 4 5
6 7 8 9