Incoming Inspection
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
2- 27
16.
Set the SDI generator for embedded audio in 2 groups starting with group 1,
and then access and set its audio channels:
a.
Press Modules, and then select the appropriate module icon, HDVG1 or
DVG1.
b.
Press Test Signals, and select Module Parameters
c.
Select Audio, and set the audio channels as follows:
H
Channel 1: 50 Hz, --35 dB
H
Channel 2: 100 Hz, --30 dB
H
Channel 3: 150 Hz, --25 dB
H
Channel 4: 200 Hz, --20 dB
H
Channel 5: 250 Hz, --15 dB
H
Channel 6: 300 Hz, --10 dB
H
Channel 7: 400 Hz, --5 dB
H
Channel 8: 500 Hz, 0 dB
17.
Connect the output of the SDI signal generator with embedded audio to the
A SDI input.
18.
Press and hold the SELECT DISPLAY 4 button to display the AUDIO
pop-up menu.
19.
In the Audio Input submenu, select Embedded.
20.
Verify that the level meter bars have a stair step pattern from –35 dB on
channel 1 to 0 dB on channel 8.
NOTE
. If level-meter bars indicate that audio is not present, set the TG2000
module number of groups to 2 even if it appears to be set already.
21.
Record Pass or Fail for Embedded Display in the test record.
22.
Bring up the audio tile pop-up and set the phase pair to 1 & 2.
23.
Go through the other phase pairs (3 & 4, 5 & 6, 7 & 8) and verify that the
phase display changes in each one.
24.
Record Pass or Fail for Embedded to Lissajous in the test record.