Specifications
1- 10
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
Table 1- 13: General characteristics (Cont.)
Characteristic
Reference information
Performance requirement
Line Select Bright-up
Selected line is indicated with a cursor in
picture and field sweeps.
Picture cursor can be disabled.
Line Select Range
Any line, numbered according to appropri-
ate standard, frame and color frame
structure.
Fields greater than 1 display the line
number in the field and the line number in
the frame or color frame.
Acoustic Noise Level, typical
Front 40, rear 39, side 38 dB(A) for stand-
ing location at 25
_
C ambient.
Front 41, rear 45, side 38 dB(A) for sitting
location at 25
_
C ambient.
Measured as per ISO7779. Unit placed
on desk, measured in operator location.
Fan speed and noise increase at higher
temperature.
Table 1- 14: Program error alarms
Characteristic
Performance requirement
Reference information
Alarm Coverage
Alarms only reported on selected input.
See Table 1--45 on page 1--34 for names
and descriptions of common alarms.
Alarm Log Depth
10,000 alarm entries.
Volatile
Alarm Log Time stamping
Time of Day, and Selectable LTC, VITC,
or ATC (Ancillary Time Code).
Alarm Notification
All alarms may be configured to be
reported to any or all of the following:
XGA display Icon or Text, Beeper, SNMP,
Ground Closure Alarm Output, Web
based User Interface, Alarm Log.
Table 1- 15: Ethernet
Characteristic
Performance requirement
Reference information
SNMP
For instrument control & feedback of
status. Complies with SNMP version 2.
IP Address Mode
Supports manual and DHCP.
Connector
RJ--45 LAN connector supporting 10/100
BaseT.
The 10/100Base--T Ethernet port connec-
tor arrangements:The RJ--45 connector
has built in LEDs. The Green LED
indicates an active connection. The
Yellow LED indicates speed. ON = 100,
OFF = 10.