Video and General Performance Verification
2- 48
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
10.
If the HD Jitter Noise Floor is outside the specification, perform HD CDR
Center and HD CDR Null calibration as described below. Adjustments for
frame rates such as 59.94 and 29.97 Hz are indicated as “1/M”; frame rates
such as 60, 50, and 30 Hz have no “1/M” indication.
a.
Press CONFIG, select Utilities > Calibration and press SEL to access the
Calibration menu.
b.
Using left and right arrow buttons, navigate to the appropriate Eye HD
CDR Center calibration case, press SEL to start, and follow the
on-screen instructions.
c.
Navigate to the appropriate Eye HD CDR Null calibration case press
SEL to start, and follow the on-screen instructions. Be sure to use the
SDI PLL Test pathological signal for this adjustment.
d.
Restart the HD Jitter Noise Floor test, and then record the new values in
the test record.
11.
Proceed to the next step if option PHY is present. Otherwise skip to step 20.
12.
Press FULL.
13.
Press <, and then select Display Type > Jitter Display.
14.
Press <, and then select Jitter1 HP Filter > 100 Hz.
15.
Press the DISPLAY SELECT 1 button briefly to close the menu.
16.
Press 5X GAIN.
17.
Measure the amplitude of the jitter waveform using the graticule or cursors.
18.
Check that the jitter amplitude is approximately 90 ps (0.13 UI), depending
on the TG2000 output jitter.
19.
Enter Pass or Fail in the test record for HD Jitter Waveform Noise.
10 second 10 Hz Jitter Bounce
20.
Press and hold the DISPLAY SELECT 1 button, and then select Jitter1 HP
Filter > 10 Hz.
21.
Watch the display for at least 10 seconds. The jitter amplitude may jump
occasionally.
22.
Check that any jump is less than 200 ps (0.297 UI).
23.
Enter Pass or Fail in the test record for HD 10 second Jitter Bounce.
24.
Turn Cursors OFF.