Video and General Performance Verification
2- 52
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
6.
Press SEL to begin this test.
7.
Press the DISPLAY SELECT 1 button.
8.
Position the display or use cursors to measure the waveform amplitude.
Check that the amplitude is between 760 mV and 840 mV.
9.
Record this level in the test record for Eye Gain.
10.
If the Eye Gain is outside the specification, perform the following adjust-
ment procedure:
a.
Press the DISPLAY SELECT 4 button to return to the Calibration menu.
b.
Use the General knob or up and down arrow buttons to adjust the
displayed amplitude of the eye waveform until it is 800 mV.
c.
Press and hold STATUS, and then select SAVE and press SEL.
d.
Repeat the Eye Gain test from step 1 and record the new values in the
test record for Eye Gain.
11.
Record this level in the test record.
12. Option PHY only.
Check that the Eye Amplitude readout in tile 2 is
between 750 mV and 850 mV.
13.
Repeat steps 3 through 12 for the SDI B input.
14.
Press the DISPLAY SELECT 4 button to return to the Calibration menu.
15.
Press and hold STATUS, and then select Exit. Press SEL to exit the
Calibration menu without saving.
This test verifies the eye signal bandwidth.
NOTE
. The level accuracy of the RF generator and the frequency response of the
cable and matching pad directly affect this check. To accurately verify this
specification with the recommended equipment, the combination of generator,
pad, and cable must be characterized as described in
Signal Source Character-
ization for Eye Signal Bandwidth
on page 2--37.
1.
Assemble the signal source setup as characterized in Signal Source Charac-
terization for Eye Signal Bandwidth Verification (see page 2--37).
2.
Set the RF generator frequency to 100 MHz.
3.
Set the RF generator level as indicated in Table 2--3 on page 2--38.
4.
Connect the signal source output to the SDI A input.
Eye Signal Bandwidth
(WVR6100 & WVR7100
Option EYE or PHY)