![Tektronix WVR6100 Series Instructions Manual Download Page 114](http://html1.mh-extra.com/html/tektronix/wvr6100-series/wvr6100-series_instructions-manual_1078279114.webp)
Video and General Performance Verification
2- 60
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
10 second 10 Hz Jitter Bounce
18.
Press and hold the DISPLAY SELECT 1 button, and then select Jitter1 HP
Filter > 10 Hz.
19.
Watch the display for at least 10 seconds. The jitter amplitude may jump
occasionally.
20.
Check that any jump is less than 300 ps (0.08 UI).
21.
Enter Pass or Fail in the test record for SD 10 second Jitter Bounce.
This test checks the SD jitter readout accuracy.
1.
Connect the DVG1 Option S1 output (typically the second output) to the
SDI A input.
2.
Set the DVG1 to generate a 525--270 100% Color Bars signal.
3.
Press the DISPLAY SELECT 1 button to select tile 1.
4.
Press <, and then select Display Type > Eye Display.
5.
Press <, and then select Jitter1 HP Filter > 100 Hz.
6.
If the jitter readout is not displayed, press <, and then select Jitter Meter >
Meter & Readout.
7.
Adjust the DVG1 SDI module parameters as follows:
a.
Set the Jitter Freq to 9950 Hz.
b.
Set the Jitter Ampl to 11.1 ns (3.00 UI).
8.
Check that the jitter readout is between 2.65 UI and 3.35 UI.
9.
Record pass or fail in the test record for SD Jitter Readout Accuracy.
SD Jitter Readout
Accuracy
(WVR6100 Options EYE or
PHY, WVR7100 Options
SD and EYE or PHY)