![Tektronix WVR6100 Series Instructions Manual Download Page 72](http://html1.mh-extra.com/html/tektronix/wvr6100-series/wvr6100-series_instructions-manual_1078279072.webp)
Incoming Inspection
2- 18
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
Table 2- 1: Required test equipment for incoming inspection (Cont.)
Test equipment
Example
Requirements
Dolby bit-stream generator
Generate Dolby Digital, Dolby E and PCM
bit streams at 48 kHz.
AES3ID--1995/SMPTE 276M output (75
Ω
BNC)
Dolby Laboratories DM100 Bitstream
Analyzer. (Options DD and DDE only.)
SDI serial digital video test generator, with
embedded audio and composite signal
NTSC Black
Tektronix TG2000 with BG1 and additional
modules indicated below.
p
g
source
1080i 59.94 HD signals required for
WVR7000, WVR7100:
100% color bars
10-bit shallow ramp
SDI Matrix Split Field Pathological Signal
100% sweep
HDVG1 module for TG2000 (Embedded
audio needed for audio options AD, DS,
DD, DDE)
525/270 SD signals required for WVR6100:
100% color bars
10-bit shallow ramp
SDI Matrix Split Field Pathological Signal
100% sweep
DVG1 module for TG2000 (Embedded
audio needed for audio options AD, DS,
DD, DDE)
Composite signals required for Option
CPS:
NTSC SMPTE bars
AVG1 module for TG2000
AES Audio Signal Generator
48kHz, 24 bit word length signals
Rohde & Schwarz UPL06, Tektronix AM700
and AM70.
AES Audio Signal Analyzer
Rohde & Schwarz UPL06, Tektronix
AM700.
Analog Audio Signal Generator
Rohde & Schwarz UPL06, Tektronix AM700
and AM70.
Video Test Signals
SDI 525 10-bit shallow ramp
SDI 525 100% sweep
NTSC black
NTSC SMPTE bars
Provided by Tektronix TG2000 as specified
above
LTC generator
Horita TG--50 or Adrienne AEC--Box--28
Voltmeter
Fluke 87 or equivalent
15-pin Dsub male connector and cable
Used to mate with the ground closure port
See Figure 2--1 on page 2--45 for wiring
diagram of cable
Computer and ethernet cable
Used to test Ethernet connection
Generic equipment