Test Records
2- 8
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
SD Eye Performance Test Record (WVR7100 Options SD and EYE or PHY, WVR6100 Option EYE or
PHY)
Instrument Serial Number:
Certificate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
Performance Test
Minimum
Incoming
Outgoing
Maximum
SD Cable Meter (0 m), SDIA
0 m
17 m
SD Cable Meter (100 m), SDIA
76 m
127 m
SD Cable Meter (0 m), SDIB
0 m
17 m
SD Cable Meter (100 m), SDIB
76 m
127 m
SD Jitter Readout Noise
Pass
SD Jitter Waveform Noise (option PHY only)
Pass
SD 10 second Jitter Bounce
Pass
SD Jitter Readout Accuracy
Pass
SD Jitter Waveform Accuracy (option PHY only)
Pass
SD Jitter Vibration
Pass
SD Recovered Sine Wave Fidelity
Pass