Incoming Inspection
2- 28
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
1.
Press FACTORY to restore the factory preset.
2.
If you have just finished testing the Digital Audio Inputs, proceed to step 3,
otherwise do steps 15--20 of the Digital Audio Input test.
3.
Press CONFIG to display the CONFIG menu.
4.
Select Audio Inputs/Outputs.
5.
Under the AES B submenu, configure the port as an Output.
6.
Press CONFIG to close the menu.
7.
Connect the AES B 1-2 output to the Audio Signal Analyzer input.
8.
Verify on the analyzer that the signal levels are at –35 dB for the left channel
and –30 dB for the right.
9.
Connect the AES B 3-4 output to the Audio Signal Analyzer input.
10.
Verify on the analyzer that the signal levels are at –25 dB for the left channel
and –20 dB for the right.
11.
Connect the AES B 5-6 output to the Audio Signal Analyzer input.
12.
Verify on the analyzer that the signal levels are at –15 dB for the left channel
and –10 dB for the right.
13.
Connect the AES B 7-8 output to the Audio Signal Analyzer input.
14.
Verify on the analyzer that the signal levels are at –5 dB for the left channel
and 0 dB for the right.
15.
Record Pass or Fail for each output in the test record.
1.
Press FACTORY to restore the factory preset.
2.
Press and hold the SELECT DISPLAY 4 button to display the AUDIO
pop-up menu.
3.
In the Audio Input submenu, select AES A.
4.
Press the SELECT DISPLAY 4 button again to dismiss the pop-up.
5.
Set the Dolby generator for a Dolby D output using the following steps:
a.
Press Gen to display the Gen Stream Sel message
b.
Press Up/Down buttons to select a Dolby D stream as indicated by the
leading “D” in the bit stream name.
c.
Press Enter to activate the selected signal.
Digital Audio Output
(Options DA, DG, DS, AD,
DD, & DDE only)
Dolby Decode (Options
DD & DDE only)