Video and General Performance Verification
2- 62
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
5.
Press < to navigate to the end of the list of Calibration Cases, highlighting
Eye Recovered Clock Test. Press SEL to begin this test.
6.
Press the DISPLAY SELECT 1 button, and then press FULL to view the
sine wave. Check that it is smooth and undistorted. There should be one
cycle displayed.
7.
Press GAIN 5X and VAR. Use the GENERAL knob to adjust the gain to
x10.
8.
Check the waveform. There should be no discontinuities or jagged edges.
Use the VERTICAL knob to view the entire waveform.
9.
Record pass or fail in the test record.
10.
Press the DISPLAY SELECT 4 button to return to the Calibration menu.
11.
Press and hold STATUS, and then select Exit. Press SEL to exit the
Calibration menu without saving.