Incoming Inspection
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
2- 23
1.
Press FACTORY to restore the factory preset.
2.
Connect an SDI 10-bit shallow-ramp signal to the SDI A input. Use a signal
type appropriate to the unit under test:
H
WVR7000 and WVR7100: 1080i 59.94 10-bit shallow ramp from
HDVG1
H
WVR6100: 525/270 10-bit shallow ramp from DVG1
3.
Terminate the SDI loop-through connector.
4.
Apply an NTSC signal to the External Reference input from the Analog
Signal Generator. Install a termination on the loopthrough.
5.
Select EXT REF on the waveform rasterizer front panel.
6.
The status bar in the lower left-hand corner of the display should display
Ref: NTSC.
7.
Press the DISPLAY SELECT 4 button, then WFM, and then FIELD.
8.
Verify that both WFM tiles are stable.
9.
Record Pass or Fail for NTSC Lock in the test record.
10.
Remove the NTSC signal from the reference input.
11.
Both WFM tiles should “unlock” and scroll.
12.
The status bar in the lower left-hand corner of the display should display
Ref: Ext. Missing.
13.
Record Pass or Fail for Ref Missing in the test record.
1.
Press FACTORY to restore factory preset.
2.
Connect an SDI color bars signal to the SDI A input. Use a signal type
appropriate to the unit under test:
H
WVR7000 and WVR7100: 1080i 59.04 color bars from HDVG1
H
WVR6100: 525/270 color bars from DVG1
3.
Terminate the SDI loop-through connector.
4.
Press EYE (if no EYE button, press MEAS then select Display Type > Eye
Display).
5.
Check for the presence of a stable eye diagram in tile 1. (Actual wave shape
depends strongly on the generator and terminator.)
6.
Record Pass or Fail for Eye Waveform in the test record.
External Reference
Eye
(Options EYE and PHY
Only)