WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
2- 17
Incoming Inspection
This section contains functional and operational checks appropriate to an
incoming inspection.
The waveform rasterizer must have been operating for a warm-up period of at
least 20 minutes, and must be operating at an ambient temperature as listed in
Table 1--20 on page 1--12.
Use the following procedures to check the basic functionality of WVR6100,
WVR7000, and WVR7100 Waveform Rasterizers. The checks are arranged by
model and option so that you can choose the sections that are appropriate for
your instrument. The last two sections are for less critical waveform rasterizer
features: the ground closure and Ethernet ports. You need only test these if you
intend to use them. In general, you should test in the order presented, since later
tests might depend on items checked in the earlier tests.
Required Equipment
The following equipment is required to perform the incoming inspection
procedures.
Table 2- 1: Required test equipment for incoming inspection
Test equipment
Requirements
Example
XGA Monitor
Computer monitor capable of 1024 x 768 x
60Hz scan rate
75
Ω
coaxial cables
(3 required)
General purpose digital video
Male-to-male BNC connectors 1 or 2
meters long
Belden 8281.
Tektronix part number:
012-0159-00 or 012-0159-01
75
Ω
Terminator for Analog Video
Male connector, precision
Tektronix part number
011-0102-03
Analog audio breakout cable (required for
Audio Option DA only)
DB37 to XLR I/O
Tektronix part number
012-1658-01
Analog audio breakout cable (required for
Audio Options AD, DD, DDE only)
DB62 to XLR I/O
Tektronix part number
012-1688-00