Video and General Performance Verification
2- 46
WVR6100, WVR7000, and WVR7100 Specifications and Performance Verification
Apply a square wave to the LTC input and use variable gain to normalize the
display height. The variable gain factor indicates the input gain error.
1.
Press FACTORY on the waveform rasterizer.
2.
Connect the RCA connector from the custom cable to the output of the
Timecode generator. Connect the 15-pin connector on the custom cable to
the REMOTE connector on the rear of the waveform rasterizer.
3.
Unplug the LTC generator so it is turned off. It needs to stay connected to
the DUT to terminate the input.
4.
Connect the 10 kHz square wave source to the BNC on the remote cable. Set
the square wave amplitude to 5 V. If necessary, use an oscilloscope to verify
that the amplitude is accurate.
5.
Press OTHER on the waveform rasterizer front panel to display the LTC
waveform.
6.
Press FULL to change the waveform rasterizer display to full-screen mode.
7.
Press VAR to activate the variable gain function.
8.
Adjust the gain and vertical position until the displayed waveform is five
divisions high.
The variable gain readout indicates the error in the input circuit. A gain of 1.03
indicates the input gain error is –3%. Record the gain error in the test record.
This test uses a short (1 to 2 meters) length of Belden 8281 cable and a long
(100 meters) length of Belden 8281 cable to check the Cable Meter. A cable
clone may be used instead of the long cable.
1.
Press FACTORY.
2.
Press and hold STATUS, and then select Display Type > SDI Status.
3.
Using the short cable, connect an HDVG1 output to the waveform rasterizer
SDI A input. Terminate the other SDI A connector with a 75
Ω
SDI
terminator.
4.
Set the HDVG1 to generate a 1080i 59.94 75% Color Bars signal.
5.
Check that the cable length reading is between 0 m and 17 m.
6.
Record the result in the test record.
7.
Replace the short cable with the long one, or set the cable clone to 100 m.
8.
Check that the cable length reading is between 76 m and 127 m.
9.
Record the result in the test record.
LTC Waveform Accuracy
HD Cable Meter
(WVR7100 Option EYE
or PHY)