Maintenance
WFM 601 - Service Manual
6–5
Corrective Maintenance
The following procedure is designed to assist in isolating problems, which in
turn expedites repairs and minimizes down time.
1. Ensure that the malfunction exists in the instrument. This is done by making
sure that the instrument is operating as intended by Tektronix (see Operating
Instructions), and by checking that a malfunction has not occurred up stream
from the waveform monitor.
2. Determine and evaluate all trouble symptoms. This is accomplished by
isolating the problem to a general area such as an assembly. The block
diagram is a valuable aid in signal tracing and circuit isolation.
CAUTION.
Use extreme care when probing with meter leads or probes, because of
the high component density and limited access within the instrument. The
inadvertent movement of leads or a probe could cause a short circuit or transient
voltages capable of destroying components.
3. Determine the nature of the problem. Attempt to make the determination of
whether the instrument is out of calibration or if there has been a component
failure. Once the type of failure has been determined, proceed on to identify
the functional area most likely at fault.
4. Visually inspect the suspect assembly for obvious defects. Most commonly
these will be broken or loose components, improperly seated components,
overheated or burned components, chafed insulation, etc. Repair or replace
all obvious defects.
5. Use successive electrical checks to locate the source of the problem. The
primary tool for problem isolation is the oscilloscope. Use the Performance
Check Procedure to determine if a circuit is operating within specifications.
At times it may be necessary to change a calibration adjustment to determine
if a circuit is operational, but since this can destroy instrument calibration,
care should be exercised. Before changing an adjustment, note its position
so that it can be returned to its original setting.
6. Determine the extent of the repair. If the necessary repair is complex, it may
be advisable to contact your local Tektronix field office or representative
before continuing. If the repair is minor, such as replacing a component, see
the parts list for replacement information. Removal and replacement
procedures for the assemblies can be found under Corrective Maintenance.
General Troubleshooting
Techniques
Summary of Contents for WFM601
Page 4: ...The WFM601 Serial Digital Component Monitor...
Page 10: ...Contents iv WFM 601 Service Manual Revised Nov 1994...
Page 14: ...Contents viii WFM 601 Service Manual Revised Nov 1994...
Page 19: ......
Page 32: ......
Page 50: ...Installation 2 18 WFM 601 Service Manual...
Page 57: ......
Page 87: ...Theory of Operation 3 30 WFM 601 Service Manual...
Page 88: ......
Page 110: ...Performance Verification 4 22 WFM 601 Service Manual...
Page 111: ......
Page 132: ......
Page 168: ...Maintenance 6 36 WFM 601 Service Manual...
Page 169: ......
Page 175: ...Options 7 6 WFM 601 Service Manual...
Page 176: ......
Page 177: ......
Page 226: ......
Page 233: ...WFM 601 SERIAL COMPONENT MONITOR BNC INPUT BOARDS 1...
Page 237: ...ECL TO TTL CONVERTER LINE BUFFER RAM WFM 601 SERIAL COMPONENT MONITOR COPROCESSOR 3...
Page 245: ...Y FILTER Pb FILTER Pr FILTER WFM 601 SERIAL COMPONENT MONITOR Y Pb Pr RECONSTRUCTION FILTERS 7...
Page 247: ...WFM 601 SERIAL COMPONENT MONITOR TRANSCODERS PIX MONITOR OUTPUTS 8...
Page 249: ...0 0 0 0 WFM 601 SERIAL COMPONENT MONITOR LIGHTNING VECTOR BOWTIE SWITCHING 9...
Page 251: ...WFM 601 SERIAL COMPONENT MONITOR CONTROL DACULATOR 10...
Page 259: ...ROM MICROPROCESSOR NOVRAM RAM FLASH EPROM BUFFERS RESET ABORT WFM 601 SERIAL COMPONENT MONITOR...
Page 265: ......
Page 271: ...WFM 601 SERIAL COMPONENT MONITOR FRONT PANEL 20...
Page 282: ...WFM 601 FIG 1 EXPLODED VIEW A1 A2 A3 A4 A4A1 A5 A6 A7 A8...
Page 290: ...Glossary Glossary 8 WFM601 Serial Component Monitor...