PERFORMANCE CHECK/CALIBRATION
Change information, if any, affecting this section will be found at the rear of this manual.
Introduction
To assure instrument accuracy, check the calibration o f
the Type 422 every 1000 hours o f operation, or every six
months if used infrequently. Before complete calibration,
thoroughly clean and inspect this instrument as outlined in
the Maintenance section.
As an aid to the calibration of the instrument, a
Short-Form Procedure is given prior to the complete
procedure. To facilitate instrument calibration fo r the
experienced calibrator, the Short-Form Procedure lists the
calibration adjustments necessary fo r each step and the
applicable tolerances. This procedure also includes the step
number and title as listed in the complete Performance
Check/Calibration Procedure and the page number on
which each step begins. Therefore, the Short-Form Pro
cedure can be used as an index to locate a step in the
complete procedure. Another feature o f the Short-Form
Procedure is the spaces provided to record performance
data or to check o ff steps as they are completed. This pro
cedure can be reproduced and used as a permanent record
of instrument calibration.
The complete Performance Check/Calibration Procedure
can be used to check instrument performance w itho ut
removing the covers or making internal adjustments by
performing all portions except the ADJUST- part of a step.
Screwdriver adjustments which are accessible w ithout
removing the covers are adjusted as part of the performance
check procedure. A note titled PERFORMANCE CHECK
ONLY gives instructions which are applicable only to the
performance check procedure and if necessary, lists the
next applicable step fo r the performance check procedure.
Completion o f each step in the complete Performance
Check/Calibration Procedure insures that this instrument
meets the electrical specifications given in Section 1. Where
possible, instrument performance is checked before an
adjustment is made. For best overall instrument perform
ance when performing a complete calibration procedure,
make each adjustment to the exact setting even if the
CHECK— is w ithin the allowable tolerance.
NOTE
All waveforms shown in this procedure were taken
with a Tektronix Oscilloscope Camera System.
Limits, tolerances and waveforms in this procedure
are given as calibration guides and should not be
interpreted as instrument specifications except as
specified in Section 1.
A partial calibration is often desirable after replacing
components, or to touch up the adjustment o f a portion of
the instrument between major recalibrations. To check or
adjust only part o f the instrument, set the controls as given
under Preliminary Control Settings and start w ith the
nearest test equipment picture preceding the desired
portion. If any controls need to be changed from the
preliminary settings fo r this portion o f the calibration
procedure, they are listed under the heading Partial
Procedure following the equipment required picture. To
prevent unnecessary recalibration o f other parts o f the
instrument, readjust only if the tolerance given in the
CHECK— part of the step is not met. If readjustment is
necessary, also check the calibration o f any steps listed in
the INTERACTION— part o f the step.
TEST EQUIPM ENT R EQ U IR ED
General
The follow ing test equipment and accessories, or its
equivalent, is required fo r complete calibration o f the Type
422. Specifications given are the minimum necessary fo r
accurate calibration. Therefore, some o f the recommended
equipment may have specifications which exceed those
given. A ll test equipment is assumed to be correctly
calibrated and operating w ithin the given specifications. If
equipment is substituted, it must meet or exceed the
specifications o f the recommended equipment.
Special Tektronix calibration fixtures are used in this
procedure only where they facilitate calibration. These
special calibration fixtures are available from T ektronix,
Inc. Order by part number through your local T ektronix
Field Office or representative.
Test Equipment
1. Variable DC power supply1. Voltage range, at least
+11 to +35 volts; current capability, at least 2.5 amperes;
output voltage monitored w ithin 3%. For example, Trygon
Model HR40-5B.
2. Precision DC voltm eter1. Accuracy, w ithin 0.1%;
resolution, 50 microvolts; range, zero to 100 volts. For
example, Fluke Model 825A Differential DC Voltm eter.
3. Test oscilloscope. Bandwidth, DC to five megahertz;
minimum deflection factor, one m illivolt/division; accur
acy,
w ithin 3%. Tektronix Type 422 Oscilloscope
recommended (use X I 0 gain feature to obtain one m illivolt
minimum deflection factor).
1 N ot required for performance check only.
5-1
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......