Specification—Type 422 AC-DC
Characteristic
Performance
Battery
Batteries
20 size D nickel-cadmium cells
supplied in battery pack, Tek
tro n ix Part No. 016-0066-02
(optional accessory).
Charge tim e (Power
Mode switch set to
CHARGE BATT
115V AC or 230V
AC)
A t least 16 hours to reach fu ll
capacity.
Operating tim e1 (bat
teries charged and
operated at +20° C
to +30° C)
SCALE ILLUM
o ff (full count
erclockwise)
A t least five hours
SCALE ILLUM
at maximum
(fully clockwise)
A t least 4.5 hours.
TABLE 1-2
ENVIRONMENTAL
NOTE
This instrument w ill meet the electrical characteristics
given in Table 1-1 over the follow ing environmental
limits. Complete details on environmental test pro
cedures, including failure criteria, etc., can be obtain
ed from Tektronix, Inc. Contact your local T ektronix
Field Office or representative.
Temperature
W ithout batteries
Operating
- 1 5 ° C to +55° C
Non-operating
- 5 5 ° C to +75° C.
With batteries
Operating
- 1 5 ° C to +40° C.
Non-operating
- 4 0 ° C to +60° C.
A ltitu de
Operating
15,000 feet maximum.
Non-operating
Test lim it 50,000 feet.
Characteristic
Performance
Hum idity
Non-operating
Five cycles (120 hours) o f Mil-
Std-202C, Method 106B. O m it
freezing and vibration. A llo w 24
hour post-test drying period at
+25°C ±5 °C and 30% to 80%
relative hum idity. L im it maxi
mum temperature to +60° C if
tested w ith batteries.
Vibration
Operating and non
operating
15 minutes along each o f the
three major axes at a total dis
placement o f 0.025-inch peak to
peak (3.9 g at 55 c/s) w ith fre
quency varied from 10-55-10 in
one minute cycles. Hold at any
resonant point, or if none, at 55
c/s fo r three minutes on each
axis.
Shock
Operating
Two guillotine-type shocks o f 30
g, one-half sine, 11-millisecond
duration each direction along
each major axis. Total o f 12
shocks.
Non-operating
One guillotine-type shock o f 60
g, one-half sine, 11-millisecond
duration in each direction along
each major axis. Total o f six
shocks.
Electromagnetic Inter
ference (EMI) As Test
ed in Mil-I-6181D
Radiated interference
Interference radiated from the
instrument under test w ithin the
given limits from 150 kilohertz
to 1000 megahertz (w ith mesh
CRT filte r installed).
Conducted interfer
ence
Interference conducted out o f
th e in s tru m e n t u n d e r test
through the power cord w ithin
the given limits from 150 kilo
hertz to 25 megahertz.
Transportation
Packaged instrument
Qualifies under National Safe
Transit Commitee test procedure
1 A, Category II.
loperating time applies to new batteries. Ampere-hour capacity
may be reduced by as much as 20% after one year.
1-5
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......