Operating Instructions—Type 422 AC-DC
hertz are attenuated. In general, AC coupling can be used
fo r most applications. However, if the signal contains un
wanted low-frequency signals or if the sweep is to be trig
gered at a low-repetition rate or DC level, one o f the re
maining TRIGGERING Coupling switch positions w ill pro
vide a better display.
The triggering point in the AC position depends upon
the average voltage level of the trigger signals. If the trigger
signals occur at random, the average voltage level w ill vary,
causing the triggering point to vary also. This shift o f the
triggering point may be enough so it is impossible to main
tain a stable display. In such cases, use DC coupling.
AC LF REJ.
In the AC LF REJ position, DC is rejected
and low-frequency signals below about 50 kilohertz are at
tenuated. Therefore, the sweep is triggered only by the
higher-frequency components o f the trigger signal. This
position is particularly useful fo r providing stable triggering
if the trigger signal contains line-frequency components.
Also, in the A L T position of the vertical Mode switch the
AC LF REJ position provides the best display at fast sweep
rates when comparing tw o unrelated signals (TRIGGERING
Coupling switch set to CH 1 & 2).
DC.
DC coupling can be used to provide stable trigger
ing w ith low-frequency signals which would be attenuated
in the AC position, or w ith low repetition rate signals. It
can also be used to trigger the sweep when the trigger signal
reaches a DC level selected by the setting o f the LEVEL
control. When using internal triggering, the setting o f the
Channel 1 and 2 POSITION controls affects the DC trigger
ing point (CH 1 & 2 position only).
DC trigger coupling should not be used in the A L T dual-
trace mode if the TRIGGERING Source switch is set to CH
1 St 2. If used, the sweep w ill trigger on the DC level of one
trace and then either lock out completely or free run on the
other trace. Correct DC triggering fo r this mode can be
obtained w ith the TRIGGERING Source switch set to CH
1.
Trigger Slope
The TRIGGERING SLOPE switch determines whether
the trigger circuit responds to the positive-going or
negative-going portion o f the trigger signal. When the
SLOPE switch is in the positive-going
(J~)
position, the
display starts with the positive-going portion o f the
waveform; in the negative-going (~\J position, the display
starts w ith the negative-going portion o f the waveform (see
Fig. 2-11). When several cycles of a signal appear in the
display, the setting o f the SLOPE switch is often
unimportant. However, if only a certain portion of a cycle
is to be displayed, the setting of the SLOPE switch is
important to provide a display which starts on the desired
slope of the input signal.
Trigger
Level
The TRIGGERING LEVEL control determines the vo lt
age level on the triggering waveform at which the sweep is
triggered. When the LEVEL control is set in the + region,
the trigger circuit responds at a more positive point on the
trigger signal. In the — region, the trigger circuit responds at
a more negative point on the trigger signal. Fig. 2-11 illus
trates this effect w ith different settings o f the SLOPE
switch.
To set the LEVEL control, first select the TRIGGER
ING source, coupling and slope. Then, set the LEVEL con
trol fu lly counterclockwise and rotate it clockwise until the
display starts at the desired point.
Trigger Mode
General.
The TRIGGERING MODE switch is actuated
by the LEVEL control. When the LEVEL control is pushed
in, the TRIGGERING Mode switch is in the AUTO position
and when the LEVEL control is pulled out, the TRIGGER
ING Mode switch is in the NORM position.
AUTO.
The AUTO position o f the TRIGGERING Mode
switch provides a stable display when the LEVEL control is
correctly set (see Trigger Level in this section) and an
adequate trigger signal is available.
When the trigger repetition rate is less than about 20
hertz, or in the absence o f an adequate trigger signal, the
Sweep Generator free runs at the sweep rate selected by the
T IM E /D IV switch to produce a reference trace. When an
adequate trigger signal is again applied, the free-running
condition ends and the Sweep Generator is triggered to
produce a stable display (with correct LEVEL control set
ting).
NORM.
Operation in the NORM position o f the T R IG
GERING Mode switch is the same as in the AUTO position
when a trigger signal is applied. However, when no trigger
signal is present, the Sweep Generator remains o ff and there
is no display.
Use the NORM mode to display signals w ith repetition
rates below about 20 hertz. This mode provides an indica
tion of an adequate trigger signal as well as the correctness
o f trigger control settings, since there is no display w itho ut
proper triggering.
Horizontal
Sweep
Rate
The T IM E /D IV switch provides 19 calibrated sweep
rates ranging from 0.5 microsecond to 0.5 second/division.
The V A R IA B LE control provides continuously variable
sweep rates between the settings o f the T IM E /D IV switch.
Whenever the UNCAL light is on, the sweep rate is uncali
brated. The light is o ff when the VA R IA B LE control is set
to the CAL position.
When making tim e measurements from the graticule, the
area between the second and tenth vertical lines o f the
graticule provides the most linear time measurement (see
Fig. 2-12). Therefore, the firs t and last divisions o f the
display should not be used when making accurate tim e
measurements. Position the start of the tim ing area to the
second vertical line and adjust the T IM E /D IV switch so the
2-19
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......