Operating Instructions—Type 422 AC-DC
Fig. 2-5. Securing the power cable to the power supply for remote
operation.
Press in here
on feofh
sides o f
the instrument to
position handle.
Fig. 2-6. Handle positioned to provide a stand for the instrument.
OPERATING TEM PERATURE
General
The Type 422 w ith AC-DC Power Supply requires very
little air circulation fo r proper operation. Thermal cutouts
in the indicator and the power supply provide thermal pro
tection and disconnect the instrument power if the inter
nal temperature exceeds a safe operating level. Power is
automatically restored when the temperature returns to a
safe level. The Type 422 w ith AC-DC Power Supply can be
operated where the ambient air temperature is between
—15°C and +55°C (+40°C maximum w ith battery pack
installed). This instrument can be stored in ambient temper
atures between ~55°C and +75°C (—40°C to +60°C w ith
batteries). A fter storage at temperatures beyond the opera
ting limits, allow the chassis to come w ithin the operating
limits before power is applied.
Operating Position
The handle of the Type 422 can be positioned fo r carry
ing or as a tilt-stand fo r the instrument. To position the
handle, press in at both pivot points (see Fig. 2-6) and turn
the handle to the desired position. Several positions are
provided fo r convenient carrying or viewing. The instru
ment can also be set on the rear-panel feet either fo r opera
tion or storage.
CONTROLS A N D CONNECTORS
General
A brief description of the function and operation of the
controls and connectors of the Type 422 follows. Fig. 2-7
shows the fro n t panel o f this instrument More detailed
inform ation is given in this section under General Operating
Information.
Cathode-Ray Tube (CRT)
INTENSITY
Controls brightness of display.
FOCUS
Provides adjustment fo r optim um dis
play definition.
ASTIGMATISM
Used in conjunction w ith the FOCUS
control to obtain a well-defined dis-
play.
SCALE ILLU M
Controls graticule illum ination.
Vertical (both channels if applicable)
VO LTS/D IV
Selects vertical deflection factor in
1-2-5 sequence (VAR IA BLE control
must be in CAL position fo r indicated
deflection).
V A R IA B LE
UNCAL
POSITION
GAIN
Provides variable uncalibrated deflec
tion factors between the calibrated set
tings o f the VO LTS/D IV switch.
L ig h t in d ic a te s that V A R IA B LE
V O LTS/D IV control is not in the CAL
position.
Controls vertical position o f the trace.
Screwdriver adjustment to set gain of
the vertical input amplifier.
m
2-7
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......