![Tektronix 422 Instruction Manual Download Page 114](http://html1.mh-extra.com/html/tektronix/422/422_instruction-manual_1078891114.webp)
Performance Check/Calibration—Type 422 AC-DC
___ !___
(A)
Expansion
-
IB)
(C)
Compression
Fig. 5-16. Typical CRT display showing acceptable compression and
expansion. (A) Expansion, (B) correct deflection at center of
graticule, (C) expansion.
e. CHECK—Compression or expansion not to exceed
0.15 division (see Fig. 5-16).
f. Position the bottom o f the display to the bottom
horizontal line o f the graticule.
g. CHECK—Compression or expansion not to exceed
0.15 division (see Fig. 5-16).
h. Set the vertical Mode switch to CH 2.
i. Position the display to the center of the graticule w ith
the Channel 2 POSITION control.
j. Set the Channel 2 V A R IAB LE VO LTS/D IV control
fo r exactly tw o divisions o f deflection (see Fig. 5-16B).
k. Position the top o f the display to the top horizontal
line of the graticule.
l
. CHECK—Compression or expansion not to exceed
0.15 division (see Fig. 5-16).
m. Position the bottom o f the display to the bottom
horizontal line of the graticule.
n. CHECK—Compression or expansion not to exceed
0.15 division (see Fig. 5-16).
22. Check Channel 1 and 2 Input Coupling Switch
Operation
a. Return both V A R IA B LE VO LTS /D IV controls to
CAL.
b. Position display w ith the Channel 2 POSITION
control so the bottom o f the square wave is at the center
horizontal line.
c. Set the Channel 2 Input Coupling switch to AC.
d<. CHECK—CRT display centered about center horizon
tal line.
e. Set the Channel 2 Input Coupling switch to GND.
f. CHECK—CRT display fo r a straight line near the cen
ter horizontal line.
g. Set the vertical Mode switch to CH 1.
h. Position display w ith the Channel 1 POSITION con
tro l so the bottom of the square wave is at the center
horizontal line.
i. Set the Channel 1 Input Coupling switch to AC.
j. CHECK—CRT display centered about center horizon
tal line.
k. Set the Channel 1 Input Coupling switch to GND.
l
. CHECK—CRT display fo r a straight line near the
center horizontal line.
m. Disconnect all test equipment.
23. Check Trace Shift Due to Input Current
a. Set both VO LTS/D IV switches to .01.
b.
Position the trace to the center horizontal line with
the Channel 1 POSITION control.
5-20
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......