![Tektronix 422 Instruction Manual Download Page 122](http://html1.mh-extra.com/html/tektronix/422/422_instruction-manual_1078891122.webp)
Performance Check/Calibration—Type 422 AC-DC
n. W ithout changing the output amplitude, set the
constant-amplitude generator output frequency to five
megahertz.
o. CHECK—Stable CRT display can be obtained with
the TRIGGERING Coupling switch set to AC, AC LF REJ
and DC (LEVEL control may be adjusted as necessary to
obtain a stable display).
p. Set the TRIGGERING Mode switch to NORM (pulled
out).
q. CHECK—Stable CRT display can be obtained with
the TRIGGERING Coupling switch set to AC, AC LF REJ
and DC (LEVEL control may be adjusted as necessary to
obtain a stable display).
r. Set the Channel 1 VO LTS/DIV switch to .1.
s. Set the constant-amplitude generator fo r a six-division
display (0.6 volt) at its reference frequency (50 kilohertz).
t. W ithout changing the output amplitude, set the
constant-amplitude generator to 15 megahertz.
u. Pull the X I 0 MAG switch
v. CHECK—Stable CRT display can be obtained w ith
the TRIGGERING Coupling switch set to AC, AC LF REJ
and DC (LEVEL control may be adjusted as necessary to
obtain a stable display).
w. Set the TRIGGERING Mode switch to AUTO
(pushed in).
x. CHECK—Stable CRT display can be obtained w ith
the TRIGGERING Coupling switch set to AC, AC LF REJ
and DC (LE VEL control may be adjusted as necessary to
obtain a stable display).
y. Disconnect all test equipment.
37. Check Low-Frequency Triggering Operation
a. Change the following control settings:
b. Connect the low-frequency constant-amplitude sine-
wave generator to the INPUT 1 connector through the BNC
to dual binding post adapter, 42-inch 50-ohm BNC cable,
50-ohm BNC term ination and BNC T connector. Connect
the o utput o f the BNC T connector to the TRIG IN con
nector w ith an 18-inch 50-ohm BNC cable.
c. Set the low-frequency generator for a 2.5-division dis
play (0.125 volt) at one kilohertz.
d. W ithout changing the output amplitude, set the
constant-amplitude generator output frequency to 50 hertz.
e. CHECK—Stable CRT display can be obtained with
TRIGGERING Coupling switch set to AC and DC (LEVEL
control may be adjusted as necessary to obtain a stable
display).
f. Set the TRIGGERING Mode switch to NORM (pulled
out).
g. CHECK—Stable CRT display can be obtained with
the TRIGGERING Coupling switch set to AC and DC
(LEVEL control may be adjusted as necessary to obtain a
stable display).
h. Set the TRIGGERING Source switch to CH 1 & 2.
i. Set the low-frequency generator for a 0.2-division dis
play at 50 hertz.
j. CHECK—Stable CRT display can be obtained w ith the
TRIGGERING Coupling switch set to AC and DC and the
TRIGGERING Source switch set to CH 1 and CH 1 & 2
(LEVEL control may be adjusted as necessary to obtain a
stable display).
k. Set the TRIGGERING Mode switch to AUTO
(pushed in).
l
. CHECK—Stable CRT display can be obtained w ith the
TRIGGERING Coupling switch set to AC and DC and the
TRIGGERING Source switch set to CH 1 and CH 1 & 2
(LEVEL control may be adjusted as necessary to obtain a
stable display).
38. Check Low-Frequency Reject Operation
a. Change the follow ing control settings:
TRIGGERING Coupling
AC LF REJ
TRIGGERING Mode
NORM
T IM E /D IV
.1 ms
b. Set the low-frequency generator fo r a 0.2-division dis
play at 50 kilohertz.
c.
CHECK—Stable CRT display can be obtained with
the TRIGGERING Source switch set to CH 1 & 2 (LEVEL
control may be adjusted as necessary to obtain a stable
display).
d. Set the low-frequency generator fo r a two-division
display at 50 kilohertz.
e. W ithout changing the o utput amplitude, set the low-
frequency generator to 50 hertz.
f. Set the T IM E /D IV switch to 5 ms.
g. CHECK—Stable CRT display cannot be obtained at
any setting of the LEVEL control w ith the TRIGGERING
Source switch set to CH 1 and CH 1 & 2.
5-28
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......