Performance Check/Calibration—Type 422 AC-DC
55. Check External Horizontal Bandwidth
a. Turn the HORIZ ATTEN control fu lly clockwise.
b. Connect the low-frequency constant-amplitude sine-
wave generator to the HORIZ IN connector through the
BNC to dual binding post adapter, 42-inch 50-ohm BNC
cable and 50-ohm BNC termination.
c. Set the constant-amplitude generator fo r six divisions
of horizontal deflection at 50 kilohertz.
d. W ithout changing the output amplitude, increase the
output frequency o f the constant-amplitude generator until
the display is reduced to 4.2 divisions ( - 3 dB point).
e. CHECK—Output frequency of generator must be 500
kilohertz or higher.
f. Disconnect all test equipment.
56. Check External Blanking
a. Connect the low-frequency sine-wave generator to the
INPUT 1 connector through the BNC to dual binding post
adapter, 42-inch BNC cable, the BNC T connector and the
50-ohm BNC termination.
b. Change the following control settings:
T IM E /D IV
10 ps
X I 0 MAG
Pushed in
c. Set the sine-wave generator fo r a four-division display
(two volts positive, peak) at 50 kilohertz.
d. Adjust the LEVEL control fo r a stable display.
e. Connect the output of the BNC T connector the EXT
BLAN KING connector w ith an 18-inch BNC cable.
f. CHECK—The positive peaks o f the displayed signal
should be blanked w ith a normal INTENSITY control set
ting (see Fig. 5-31).
g. Disconnect all test equipment.
57. Check/Adjust Calibrator Amplitude
PERFORMANCE CHECK O NLY
Parts a through g of this step are not applicable to a
performance check. Proceed to part h.
a. Connect the precision DC voltmeter from pin connec
tor G on the Calibrator and Regulators board (see Fig.
5-32) to chassis ground.
b. Remove transistor Q775 (see Fig. 5-32) from its
socket.
c. CHECK—Meter reading; —2 volts ±10 m illivolts (w ith
in 0.5%).
Fig. 5-31. Typical CRT display when checking external blanking.
Q775
Fig. 5-32. Location of calibrator test points and amplitude
adjustment (Calibrator and Regulator board).
d. ADJUST—Calibrator Am plitude adjustment R780
(see Fig. 5-32) fo r exactly —2 volts.
e. Connect the precision DC voltmeter from pin connec
tor J on the Calibrator and Regulators board (see Fig. 5-32)
to chassis ground.
f. CHECK—Meter reading: —200 m illivolts ± 3 m illivolts
(within 1.5%).
g. Disconnect the voltmeter and replace Q775.
5-36
h. Connect the 1X probe to the INPUT 1 connector.
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......