Performance Check/Calibration—Type 422 AC-DC
52. Check Gate O utput Signal
Page 5-33
REQUIREMENT: 0.5 volt or greater amplitude; dura
tion at 1 ms/DIV, between 5.2 and 6.05 divisions at test
oscilloscope sweep rate o f tw o milliseconds/division.
PERFORMANCE:_____ volt am plitude;_____ division
duration.
53. Check Chopped Operation
Page 5-33
REQUIREMENT: Repetition rate, 150 kilohertz ±20%
indicated by cycle duration of 5.6 to 8.4 divisions at 1
jUs/DIV; time segment displayed from each channel, 2.0 to
4.7 microseconds; switching transients blanked out.
PERFO RM ANCE:----------cycle_d u ra tio n ;--------- micro
second tim e segment; blanking correct
; incorrect____.
54. Check External Horizontal Operation
Page 5-35
REQUIREMENT: Deflection o f 4.0 to 6.7 divisions w ith
50-volt signal (X10 MAG switch pushed in, HORIZ ATTEN
clockwise); deflection of 4.0 to 6.7 divisions w ith five-volt
signal (X10 MAG switch pulled out, HORIZ ATTEN
clockwise) w ith a 10:1 reduction in deflection factor when
HORIZ ATTEN is turned fu lly counterclockwise.
PERFORMANCE: C orrect----- ; inco rre ct--------
55. Check External Horizontal Bandwidth
Page 5-36
REQUIREMENT:
Not more than - 3 dB at 500
kilohertz.
PERFORMANCE: - 3 dB p o in t,-------megahertz.
56. Check External Blanking
Page 5-36
REQUIREMENT: Trace blanked w ith tw o-volt positive
peak signal.
PERFORMANCE: C orrect----- ; incorrect-------
57. Check/Adjust Calibrator Am plitude
Page 5-36
REQUIREMENT: - 2 volts ±10 m illivolts at C ALIBR A
TOR jack w ith Q775 removed and —200 m illivolts ±3
•millivolts internal. Check operation in CALIBRATE 4
DIVISIONS position o f Channel 1 and 2 V O LTS/D IV
switches.
PERFORMANCE: External, w ithin
m illivolts; in
ternal, w ith in ------ millivolts; CALIBRATE 4 DIVISIONS
C hannel
1 , w ith in ------- division; Channel 2. w ithin
------ division.
58. Check Calibrator Repetition Rate
Page 5-37
REQUIREMENT: Duration o f one cycle between 4.2
and 6.3 divisions at .2 ms/DIV.
PERFORMANCE:------- divisions.
59. Check Calibrator Duty Cycle
Page 5-37
REQUIREMENT: Length of positive segment o f square
wave between 4.5 and 5.5 divisions w ith one complete
cycle/10 divisions.
PERFORMANCE: Positive segment,------ divisions.
PERFORMANCE C H E C K /C A LIB R A TIO N
PROCEDURE
General
The following procedure is arranged so the Type 422 can
be calibrated w ith the least interaction of adjustments and
reconnection of equipment. A picture of the test
equipment required fo r each group o ffsteps is given to aid
in identification o f the necessary equipment. The control
settings and test equipment setup throughout this
procedure continue from the precedingiistepfs) unless noted
otherwise. The control settings can be checked at any "test
equipment required" picture by setting the controls as
given at the start o f the procedure under the heading
Preliminary Control Settings. Then make any changes listed
following the "test equipment required" picture, under the
heading Partial Procedure (also applies to partial calibration
procedure).
NOTE
Control titles which are printed on the front panel of
the Type 422 are capitalized in this procedure
(e.g.JNTENSITY). Internal adjustments are initial
capitalized only (e.g., Ch 1 Gain).
The following procedure uses the equipment listed under
Test Equipment Required. If other equipment is substi
tuted, control settings or calibration setup may need to be
altered to meet the requirements of the equipment used.
Detailed operating instructions fo r -the test equipment are
not given in this procedure. Refer to the instruction manual
fo r the test equipment if more information is required.
NOTE
This instrument should be calibrated at an ambient
temperature of +25°C ±5°C for best overall accuracy.
The performance of this instrument can be checked
at any temperature within the 0°C to +40°C range. If
the ambient temperature is outside the given range,
see Section 1 for the applicable tolerances.
Preliminary Procedure for Performance Check
Only
1. Connect the Type 422 to a power source which meets
the voltage and frequency requirements of this instrument.
5-7
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......