![Tektronix 422 Instruction Manual Download Page 129](http://html1.mh-extra.com/html/tektronix/422/422_instruction-manual_1078891129.webp)
Performance Check/Calibration-Type 422 AC-DC
Precision
DC
voltmeter
Standard
amplitude
calibrator
42-inch
50-ohm
BNC cable
Dual
binding post
adapter
50-ohm
BNC
termination
Low-frequency
constant-amplitude
sine-wave generator
BNC T
connector
1 8-inch
BNC cable
Three-inch
screwdriver
i x
probe
Fig. 5-30. Test equipment required for steps 54 through 59.
Partial Procedure
If beginning a partial procedure w ith this step, set the
controls as given under Preliminary Control Settings.
54. Check External Horizontal Operation
a. Test equipment required fo r steps 54 through 59 is
shown in Fig. 5-30.
CAUTION
To prevent permanent damage to the CRT phosphor,
reduce the INTENSITY control setting if a halo
forms around the external horizontal display.
b. Change the follow ing control settings:
HORIZ ATTEN (LEVEL)
Clockwise
T IM E /D IV
EXT HORIZ
c. Connect the standard amplitude calibrator output
connector to the HORIZ IN (TRIG IN) connector w ith the
42-inch BNC cable.
d. Set the standard amplitude calibrator fo r a 50-volt
square-wave output.
e. Center the display (two dots) on the graticule with
the horizontal POSITION control.
f. CHECK—CRT display fo r horizontal deflection be
tween 4.0 and 6.7 divisions (10 volts/division ±25%).
g. Pull the X I 0 MAG switch.
h. Set the standard amplitude calibrator fo r a five-volt
square-wave output.
i. Center the display on the graticule w ith the horizontal
POSITION control.
j. CHECK—CRT display fo r horizontal deflection be
tween 4.0 and 6.7 divisions (one volt/division ±25%). Note
the exact horizontal deflection.
k. Set the standard amplitude calibrator fo r a 50-volt
square-wave output.
l
. Turn the HORIZ ATTEN control fu lly counterclock
wise.
m. CHECK—CRT display fo r horizontal deflection
equal to or less than noted in part j (indicates 10:1 or
greater range).
n. Disconnect all test equipment.
5-35
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......