Performance Check/Calibration—Type 422 AC-DC
Fig. 5-5. Typical test oscilloscope display of power-supply ripple
(sweep rate 20 microseconds/division).
5. Adjust Low-Voltage Indicator
a. Set the Power Mode switch (rear panel) to OPERATE
I NT BATT.
b. Connect the variable DC power supply positive
output to the bottom battery-interconnection banana plug
(see Fig. 5-6) w ith an 18-inch dual banana plug-jack patch
cord. Connect the negative output to the top battery-
interconnection banana plug with another 18-inch dual
banana plug-jack patch cord.
c. Connect the precision DC voltmeter to the battery-
interconnection banana plugs.
d. Set the variable DC power supply fo r about +35 volts
output.
e. CHECK-POWER light (front panel o f indicator) is on
and does not blink.
Fig. 5-6. (A) Location of battery-interconnection banana plugs (rear
of power supply), (B) location of low-voltage indicator adjustment
(DC Power Control board).
f. Slowly decrease the output voltage o f the variable DC
power supply.
g. CHECK—POWER light begins to blink when power
supply output voltage is 22 volts +0, —1 volt as indicated
by the precision DC voltmeter.
h. ADJUST—With the variable DC power supply set to
exactly +22 volts, set the Low-Voltage Indicator adjust
ment R1047 (see Fig. 5-6B) clockwise until the POWER
light is on steady. Then, slowly adjust this control
counterclockwise until the POWER light just begins to
blink.
i. Recheck parts d through g of this step.
j. Disconnect all test equipment.
6. Check AC Operation
a. Set the Power Mode switch (rear panel) to OPERATE
115V AC (use OPERATE 230V AC fo r 230-volt nominal
line).
b. Connect the autotransformer to an AC line voltage
source.
c. Set the autotransformer output voltage to 115 (or
230) volts.
d. Connect the AC-DC Power Supply to the autotrans
former output w ith the AC power cord.
e. CHECK—O utput voltage o f each supply w ith the
precision DC voltmeter fo r tolerances given in Table 5-1.
5-12
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......