![Tektronix 422 Instruction Manual Download Page 7](http://html1.mh-extra.com/html/tektronix/422/422_instruction-manual_1078891007.webp)
Specification—Type 422 AC-DC
Characteristic
Performance
Delay Line
Permits viewing of leading edge
of triggering signal (internal trig
gering only).
TRIGGERING
Trigger Source (select
ed by front-panel
TRIGGERING
Source switch)
Internal from both Channel 1
and 2.
Internal from Channel 1 only.
External from signal applied to
TRIG IN connector.
Trigger Coupling (select
ed by front-panel
TRIGGERING
Coupling switch)
AC (capacitive)coupled.
AC (capacitive) coupled with
a tte n u a tio n of low-frequency
signals.
DC (direct) coupled.
Trigger Slope (select
ed by front-panel
SLOPE switch)
Sweep can be triggered from the
positive-going or negative-going
portion of trigger signal.
Trigger Level
LEVEL control allows selection
o f the triggering point on the
selected slope of the trigger
signal.
Trigger Mode (select
ed by front-panel
TRIGGERING Mode
switch)
AUTO Mode fo r triggered sweep
at point selected by LEVEL con
trol when adequate trigger signal
is applied (see Trigger Sensitiv
ity); sweep free runs at sweep
ra te selected by T IM E /D IV
switch when trigger signal is in
adequate or below 20 hertz.
NORM Mode fo r triggered sweep
at point selected by LEVEL con
trol when adequate trigger signal
is applied (see Trigger Sensitiv
ity); no trace when trigger signal
is inadequate.
Trigger Sensitivity
(AUTO and NORM
Mode)
Internal
AC
0.2 division of deflection m ini
mum, 50 hertz to five mega
hertz; increasing to one division
at 15 megahertz.
AC LF REJ
0.2 division of deflection m ini
mum, 50 kilohertz to five mega
hertz; increasing to one division
at 15 megahertz. Will not trigger
on two-division or less sine wave
at 120 hertz or less.
Characteristic
Performance
DC
0.2 division of deflection m ini
mum, DC to five megahertz; in
creasing to one division at 15
megahertz.
External
AC
125 m illiv o lts m inimum, 50
hertz to five megahertz; increas
ing to 0.6 volt at 15 megahertz.
AC LF REJ
125 m illivolts minimum, 50 kilo
hertz to five megahertz; increas
ing to 0.6 volt at 15 megahertz.
Will not trigger on 1.25 volt or
less sine wave at 120 hertz or
less.
DC
125 m illivolts minimum, DC to
five megahertz; increasing to 0.6
volt at 15 megahertz.
External Trigger Input
RC characteristics
100 kilohms ±3% paralleled by
33 picofarads ±5 pF.
Maximum Input vo lt
age
250 volts DC + peak AC.
LEVEL control range
A t least +10 to —10 volts.
TIM E BASE
Sweep Rate
Calibrated range
Unmagnified
0.5 second to 0.5 microsecond/
division in 19 steps. Steps in
1-2-5 sequence.
Magnified
50 milliseconds to 0.05 micro-
second/division
in
19
steps.
Steps in 1-2-5 sequence.
Uncalibrated (vari
able) range
Continuously variable between
calibrated sweep rate settings.
E x te n d s slowest uncalibrated
sweep rate to at least 1.25
seconds/division.
Accuracy
Unmagnified
Over center eight
divisions
W ithin 3% o f indicated sweep
rate.
Over any tw o divi
sion portion w ith
in center eight
divisions
Within 4% o f indicated sweep
rate.
Magnified
Over center eight
divisions
Within 5% o f equivalent sweep
rate.
1-3
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......