Performance Check/Calibration-Type 422 AC-DC
Fig. 5-24. Typical CRT display when checking sweep length.
43. Check Sweep Length
a. Position the tenth marker to the center vertical line
w ith the horizontal POSITION control (see Fig 5-24).
b. CHECK—Sweep length between 10.4 and 12.1 d ivi
sions as shown by 0.4 to 2.1 divisions o f display to the right
o f the center vertical line (see Fig. 5-24).
44. Check Magnified Timing
a. Set the time-mark generator fo r 50-microsecond mar
kers.
Fig. 5-23. (A) Typical CRT display showing correct sweep
calibration, (B) location of Sweep Cal adjustment (Horizontal
Amplifier board).
d. ADJUST—Sweep Cal adjustment R512 (see Fig.
5-23B) for one marker each division. The second and tenth
markers must coincide exactly w ith their respective grati
cule lines (reposition the display slightly w ith the horizon
tal POSITION control if necessary).
e. Position the second marker to the second vertical
line.
b. Change the follow ing control settings:
X 10M A G
Pulled out
Horizontal POSITION
Midrange
c. CHECK—CRT display fo r one marker each division
between the second and tenth vertical lines. Marker at tenth
vertical line must be w ithin 0.4 division (within 5%) o f that
graticule line when the marker at the second vertical line is
positioned exactly.
d. Position the first 10-division portion of the total mag
nified sweep onto the viewing area w ith the horizontal
POSITION control.
f. CHECK—Fourth marker w ithin 0.08 division (within
4%) o f the fourth vertical line.
g. Position the third marker to the th ird vertical line.
h. CHECK—Fifth marker w ithin 0.08 division (within
4%) of the fo urth vertical line.
i. Continue this check fo r each two-division portion of
the sweep w ith in the center eight divisions o f the graticule.
j. IN TER AC TIO N —Check steps43 through 53.
e. CHECK—One marker each division between the sec
ond and tenth vertical lines. Marker at the tenth vertical
line must be w ithin 0.4 division (w ithin 5%) o f th at grati
cule line when the marker at the second vertical line is
positioned exactly
f. Repeat this check fo r each 10-division portion o f the
total magnified sweep length.
g. Set the horizontal POSITION control to midrange.
h. Position a marker to the second vertical line
5-30
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......