Performance Check/Calibration—Type 422 AC-DC
f. Set the autotransformer output voltage to 95 or 190)
volts.
g. CHECK—O utput voltage o f each supply w ith the
precision DC voltmeter fo r the tolerances given in Table
5-1.
h. Set the autotransformer output voltage to 137 (or
274) volts.
i. CHECK—Output voltage o f each supply w ith the
precision DC voltmeter for the tolerances given in Table
5-1.
j. Disconnect the precision DC voltmeter and use the
test oscilloscope with IX probe fo r parts m through q.
k. CHECK—Ripple o f each supply fo r AC ripple
tolerances given in Table 5-1.
l
. Set the autotransformer output voltage to 95 (or 190)
volts.
m. CHECK—Ripple o f each supply for AC ripple
tolerances given in Table 5-1.
n. Set the autotransformer output voltage to 115 (or
230) volts.
o. CHECK—Ripple of each supply fo r AC ripple
tolerances given in Table 5-1.
p. Disconnect the voltmeter.
7.
Check Battery Charger Operation
a. Turn o ff the AC power to the instrument.
b. Set the Power Mode switch (rear panel) to CHARGE
BATT 115V AC (use CHARGE BATT 230V AC for
230-volt nominal line).
c. Connect the 50-ohm 10-watt resistor to one o f the
battery terminals (see Fig. 5-6A) w ith the six-inch jumper
lead. Connect the ammeter between the remaining battery
terminal and the other side of the resistor. Position all leads
so a short-circuit condition does not exist.
d. Set the autotransformer output voltage to 95 (or 190)
volts and re-apply the AC power.
e. CHECK—Ammeter reading; 325 milliamperes m ini
mum.
NOTE
Do not leave the instrument on in this condition for
extended periods o f time as the high current through
the external resistor may cause it to overheat. If the
resistor becomes overheated, turn o ff the power and
allow it to cool before continuing.
f. Set the autotransformer output voltage to 126.5 (or
253) volts.
g. CHECK—Ammeter reading; 450 milliamperes m axi
mum.
h. Turn o ff the AC power to the instrument.
i. Replace the 50-ohm resistor w ith a 680-ohm tw o-w att
resistor.
j. Set the Power Mode switch to OPERATE 115V AC
(use OPERATE 230V AC fo r 230-volt nominal line).
k. Turn on the AC power.
l
. CHECK—Ammeter reading; 30 to 50 milliamperes.
m. Set the autotransformer output voltage to 95 (or
190) volts.
n. CHECK—Ammeter reading; 30 to 50 milliamperes.
o. Turn o ff the instrument and disconnect all test
equipment. If the AC line voltage available is not about 115
(or 230) volts, leave the autotransformer connected to the
instrument and set it fo r 115 (or 230) volts.
8. Check High-Voltage Supply
a.
With the AC power o ff, remove the plastic cover from
the CRT socket.
b. Connect
the
DC voltmeter
(VOM)
from
the
high-voltage test point (pin 3 of CRT socket, see Fig. 5-7)
to chassis ground.
c. Return the POWER switch to ON (at power supply).
d. CHECK—Meter reading; —1385 volts ±69.5 volts.
e. Disconnect the voltmeter and replace the plastic
cover.
9. Adjust Unblanking Center
a. Change the follow ing control settings:
INTENSITY
Midrange
TRIGGERING Mode
AUTO (pushed in)
b. ADJUST—Unblanking Center adjustment R869 (see
Fig. 5-8) fo r maximum trace intensity. This should occur
near the center of adjustment range.
10. Check/Adjust Trace Rotation
a.
Position the trace to the center horizontal line w ith the
Channel 1 POSITION control.
5-13
Summary of Contents for 422
Page 4: ...R TV 422 AC OC P o p p Type 422 AC DC ...
Page 64: ...Circuit Description Type 422 AC DC 3 26 Fig 3 15 Idealized DC DC Regulator circuit waveforms ...
Page 68: ...NOTES ...
Page 90: ...Maintenance Type 422 AC DC 4 22 Si Fig 4 14 High Voltage Rectifier Multiplier circuit board ...
Page 92: ...Maintenance Type 422 AC DC 4 24 Fig 4 16 DC Power Converter circuit board ...
Page 93: ...Maintenance Type 422 AC DC E 4 25 Fig 4 17 DC Power Control circuit board ...
Page 94: ...NOTES ...
Page 132: ...NOTES ...
Page 158: ......
Page 180: ...B U D C K D A G A N s N e o o o o u p T Y P E 4 Z 2 A C D C 0 2 foe ...
Page 187: ...T Y P E 4 1 2 A C D C l S W E E P T R I G G E R 3 s M 2 0 0 0 0 U P 1268 SWEEP TRIGGER ...
Page 191: ... H O R I Z O N T A L A M P L I F s N z o o o o u p H O R I Z O N T A L AMPLIFIER ...
Page 192: ... ...
Page 195: ...CRT CIRCUIT 10 3 f 8 OS M hS q u a gs u 0 z 3 U js o 0 1 j C V J C J u l Q ...
Page 198: ...FIG 1 FRONT TYPE 422 AC DC ...
Page 199: ...FIG 2 CHASSIS ...
Page 200: ...4 I TYPE 422 AC DC FIG 2 ...
Page 201: ...FIG 3 AC DC POWER SUPPLY s TYPE 422 AC DC FIG 3 ...
Page 205: ......