Appendix: Warranted Speci
fi
cations, Safety Certi
fi
cations, and Electromagnetic Compatibility
European Union
EC Council EMC Directive 89/336/EEC, amended by 93/68/EEC;
Demonstrated using:
EN 61326/A2 Electrical equipment for measurement, control, and laboratory use. Annex D
1 2
Emissions: EN 61326, Class A
Immunity
IEC 61000-4-2
IEC 61000-4-3
3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
4
IEC 61000-4-11
EN 61000-3-2
EN 61000-3-3
Australia
EMC Framework, demonstrated per Emission Standard AS/NZS 2064 (Industrial, Scienti
fi
c, and Medical
Equipment)
1
Emissions that exceed the levels required by this standard may occur when this equipment is connected to a test object.
2
Use Low-EMI Shielded cables to maintain compliance.
3
The increase in trace noise, while subjected to the test
fi
eld (3 V/m over the frequency range 80 MHz to 1 GHz with 80% amplitude modulation
at 1 kHz), is not to exceed 2 major divisions peak-to-peak. Ambient
fi
elds may induce triggering when the trigger threshold is offset less than
4 minor divisions from ground reference.
4
The increase in trace noise, while subjected to the injected 3 V test signal, is not to exceed 2 major divisions peak-to-peak. Ambient
fi
elds may
induce triggering when the trigger threshold is offset less than 1 major division from ground reference.
298
Tektronix 4000 Series Oscilloscopes User Manual
Summary of Contents for 4000 Series
Page 4: ......
Page 18: ...Preface xiv Tektronix 4000 Series Oscilloscopes User Manual ...
Page 53: ...Installation 2 Power off your oscilloscope Tektronix 4000 Series Oscilloscopes User Manual 35 ...
Page 68: ...Installation 50 Tektronix 4000 Series Oscilloscopes User Manual ...
Page 312: ...Application Examples 294 Tektronix 4000 Series Oscilloscopes User Manual ...