Appendix: Warranted Speci
fi
cations, Safety Certi
fi
cations, and Electromagnetic Compatibility
DC Balance
0.2 div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.25 div at 2 mV/div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.5 div at 1 mV/div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.2 div with the input DC-1 M
Ω
coupled and 50
Ω
terminated
0.3 div at 1 mV/div with the input DC-1 M
Ω
coupled and 50
Ω
terminated
DC gain accuracy
For 1 M
Ω
path:
±1.5%, derated at 0.100%/°C above 30 °C
±3.0% Variable Gain, derated at 0.100%/°C
above 30 °C
For 50
Ω
path:
±1.5%, derated at 0.050%/°C above 30 °C
±3.0% Variable Gain, derated at 0.050%/°C
above 30 °C
Offset accuracy
±[0.005 × | offset – position | + DC Balance]
NOTE.
Both the position and constant offset term must be converted to volts by multiplying by the
appropriate volts/div term.
Long-term sample
rate and delay time
accuracy
±5 ppm over any
≥
1 ms time interval.